Plunge Freezing of Particulate Specimens
Needed Materials (Some of the numbers correspond to labels in Fig. 1 below):
- Liquid nitrogen in a Taylor-Whorton LD4 liquid nitrogen dewar or equivalent
- Several pairs of self-closing tweezers or tweezers with locking O-rings
- Pair of fine-point, normal-closing tweezers. The user may want to tape a thin layer of
insulating material to the sides of these tweezers so that they may be more easily handled
in liquid nitrogen.
- P20 Gilson Pipetman and pipet tips
- Copper EM Grids, 400mesh (Electron Microscopy Sciences or other EM supply company) coated with a perforated carbon film.
Quantifoil grids may be used instead.
These grids should be freshly glow-discharged.
- Whatman #1 filter paper cut to rectangular strips of approximately 2 cm X 9 cm size
- 350 ml vacuum flask
- 50ml conical test tube for storing grid boxes
|
- Gatan covered grid box with cover-rod. Non-Gatan cryo, grid boxes may also be used.
- Eye protection
- Ethane gas (CP grade) with regulator, tygon tubing and a pipette tip
- Aluminum foil-lined styrofoam cooler
- Laboratory "freezer" tape
- Freeze-plunge device. Our device is in-house made but commercially-made, automated devices
such as the FEI Vitrobot or the Gatan Cryoplunge are available.
- Two transfer cups and wire support ("spiders"), one for liquid nitrogen and the other for liquid ethane. Ours
were made from test tube caps (VWR Scientific products SP16) with a holder that suspends them over the
350ml vacuum flask.
- Particulate sample for freezing
|

Fig. 1. Setup needed for plunge-freezing samples.
|
| Caution: Eye protection is a necessity. Liquified ethane can do serious damage
to the eyes. If liquid ethane comes in contact with liquid nitrogen it can violently splash. Ethane is also flammable
so it must be used in an area with adequate ventilation. No open-toed shoes are allowed when
handling cryogens and protective lab garments should also be worn. Standard practices for
handling infectious materials should be followed.
|
Preparation:
- To save a great deal of effort it
is advisable to check the quality of the specimen by preparing a
negatively-stained preparation prior to examination with
cryoelectron microscopy. In our lab, with a virus specimen of
about 50 nm diameter, we have found that the sample concentration
should be about 3-5mg/ml of protein. On a good quality
negative-stain preparation the particles would then cover the
field of view and be touching or even slightly overlapping. Of
course a specimen of much greater or smaller size would require
different concentrations which would have to be determined by
trial-and-error. The amount of solute in the buffer is also
important and a solute content of <300mM is advisable. We have not found much of a difference
when using different buffers but some materials such as glycerol or detergents may affect the
quality of the frozen prep.
- Perforated carbon support films
are preferable over standard films if the images of the sample are to be used for calculating a three-dimensional
reconstruction. This is because the amorphous
carbon of the films adds additional noise to the images. However,
if the sample concentration is somewhat low then standard support
films may be preferable because the sample has a tendency to
adhere to the carbon when allowed to sit for several minutes before blotting and plunging. In recent
years Quantifoil films have been preferred over in-house made films.
Quantifoils have holes of a standard size and periodicity, a characteristic
not easily achieved in in-house-produced perforated films. Quantifoils, however, are quite a bit
more expensive than normal grids and so they may not be the choice for inexperienced users with a
less-than refined technique. Films should be freshly prepared and not be more than a few weeks old because the hydrophilic
nature of the films changes over time which affects the ability of the sample to evenly spread over the film. Various techniques have been used to get an even
spreading. Some researchers will glow
discharge their films prior to use but we have found that too
much of the sample then adheres to the carbon instead of being suspended over
the holes. This problem has sometimes been alleviated by double blotting; that is, putting a drop of
the sample on the grid, blotting it nearly dry and then adding another drop before the final blotting and plunging. Experience
has shown that Quantifoils generally need glow discharging before use.
- The temperature and humidity in the room make a big difference in evaporation of the sample
and on how much moisture contamination occurs in the nitrogen. It is best to work in a walk-in cold cooler
at 4oC to prevent this ice contamination from forming and getting on the grids.
|
Procedures (Click on each thumbnail on the right for a larger
image. Click on the large image to close.):
- Prepare your work surface area by
placing everything that you will need within arm's reach and
removing anything that you will not need. It is preferable to
work in a chemical hood or another area with adequate air flow because of the explosive nature of
ethane.
- Check to make sure that the height of the freeze-slam
device is correct. The proper positioning of the rod is when, at the press of the foot switch, the grid is under the
level of the liquefied ethane but not so far down that the grid
might slam into any frozen ethane that may be in the cup. Most of the time the height on the freeze-slam
device need not be modified because most of our vacuum flasks are the same size. Put on
eye protection and fill both the vacuum flask and the styrofoam
cooler with liquid nitrogen. Place a conical test tube, a Gatan
grid box and cover tool, and the liquid-nitrogen transfer cup into
the cooler.
- Slowly lower the ethane
transfer cup into the nitrogen in the vacuum flask. If liquid
nitrogen bounces into the cup, empty it because it will cause
difficulty in liquefying the ethane and could cause the freezing ethane to pop and splash (Fig. 2).
- Make sure that the small black valve nearest the tygon tubing on the ethane
regulator is closed and the main valve on the ethane tank is open. You will be using the small
black valve to carefully regulate the flow of ethane into the cup. Place the tip of the ethane gas
Pasteur pipette into the bottom of the ethane transfer cup and
slowly open the small black valve. The ethane will first come out as a gas
but will soon liquefy in the bottom of the cup. Listen for the "gurgling" sound of the ethane
as it starts to liquify. Keep moving the
pipette tip around the bottom of the cup to prevent ethane from
freezing in the tip. Use extra caution whenever you add ethane
because if the pipette tip freezes or ethane is added too
quickly, liquid ethane and nitrogen could be forcefully blown out
of the cup and could cause serious eye injury. As the ethane liquefies
it may be possible to increase the flow of the gas. Allow the
transfer cup to fill up until a meniscus of liquid forms at the
top. Do not overfill the cup and allow ethane to flow over the
side and freeze. Anything frozen outside of the cups could freeze the nitrogen cup in place thereby making
it difficult to transfer the frozen grid.
- Place the liquid nitrogen transfer
cup into the vacuum flask next to the ethane cup. Do this slowly to prevent nitrogen from splashing into the
ethane. Take a pair of self-closing tweezers, pick up a grid, and secure the tweezers into place on the
plunging rod with "freezer" tape. Raise the rod so that the tip of the tweezers
is approximately 8 cm above and centered over the surface of the ethane liquid.
- With the Pipetman put about 3.5μl of
the sample onto the grid. Keeping the volume of the drop the same makes the technique more
reproducible. If the drop does not spread evenly over the grid you may need to pull it around the grid
with the pipette tip. Alternatively, you may
put the sample on the grid before attaching the tweezers to the rod.
- Take a filter paper strip and hold
it in position so that it is parallel to the plane of the
grid in front of the sample drop. Do not touch the paper
against the grid yet. Position yourself so that you are
looking through the grid and paper strip from the back side.
- Touch the filter paper
flat against the sample and the grid for several seconds (Fig. 3). Observe from the back side
how the sample liquid is wicked onto the paper. When the wicking
appears to cease, quickly hit the foot switch to plunge the grid
into the ethane. The time between the initial wicking and the
plunging is at best 2-4 sec. If it is difficult to see the
wicking a strong light may need to be placed in front of the
grid. If the wicking is done for too long of a period the grid
will be dry. Conversely, if the wicking is too short the ice will
be too thick. The paper must be held parallel and firmly against
the grid without damaging the grid. If any part of the grid pulls
away from the paper before plunging, the sample will begin to
evaporate and the sample particles will be bunched together near
the periphery of the holes in the carbon film.
- Carefully remove the tape that
holds the tweezers in place on the plunging rod and very quickly
transfer the grid to the nitrogen transfer cup. If done quickly
enough a thin layer of liquefied ethane remains on both sides of
the grid and provides a momentary insulation during the transfer.
- Pick up the nitrogen transfer cup
and the tweezers and move both to the styrofoam cooler and place
into the liquid nitrogen (Figure 4). Release the grid into the Gatan grid
box. Try not to breath onto the surface of the nitrogen because
the resultant fog will make it impossible to see the grid box.
This procedure may be done somewhat more easily by picking up the
grid box with a standard pair of tweezers and holding the box
nearer to the top of the liquid nitrogen. Care must be taken to
avoid exposing the grid to the ambient air. The Gatan grid boxes and cover rods are in limited
supply. If you plan on storing your grids in liquid nitrogen for any length of time please use the square
grid boxes instead.
- If additional grids are needed,
place the tools back into their proper positions and top off all
liquids. If the ethane has frozen re-melt it by
adding more ethane gas. Use a dry pair of self-closing tweezers for
the next grid. Make sure that all tools are dry before being cooled down. Wet tools will freeze and make the process
much more difficult. For example, it will be impossible to remove a frozen grid without damaging it if wet tweezers
are used. A hair dryer may be used to warm up and dry frozen items.
- When sufficient grids have been
frozen, screw the grid box cover rod onto the top
of the grid box and then place both into the conical tube filled
with nitrogen. The conical tube and samples can then be placed
into one of the vacuum flasks and carried to the microscope or
placed into storage in a liquid nitrogen cryo freezer.
- Close the valve on the ethane bottle and clean up the area
for the next person. Pour the liquid gases out of the dewars, transfer cups, and cooler into an appropriate place
in a well-ventilated area.
|
Fig. 2. Vacuum flask in position on the freeze-slam device. The flask is filled
with nitrogen and has the two transfer cups, the one on the left is filled with liquid nitrogen and the one on
the right with liquified ethane. Some solidified ethane may be observed in the bottom of the cup.
Fig. 3. Blotting the grid before plunging.
Fig. 4. Placing the grid in the Gatan grid-storage box.
|
|